https://www.trongkamol.com/product-page/trongkamol-starrett-mkp?lang=th
top of page
Advanced Ultrasonic Proceq Flaw Detector 100

Advanced Ultrasonic Proceq Flaw Detector 100

  • Brand: Proceq (Switzerland) 

 

  • Affordable high tech flaw detector - UT / PA / TOFD in one instrument

 

  • Excellent software and reporting 

 

  • Rugged and compact

Hardware

 

Housing

• Dimensions (HxWxD) 205 mm x 300 mm x 90 mm

  (8.1 inch x 11.8 inch x 3.5 inch)
• Weight (with battery) 3.5 kg (7.7 lb)

Display • 8.4” 800 x 600 pixel resolution
• Display Colours 260k (65535 colours for scan palettes)
• Display type TFT LCD, 450 Cd/m2, with 2% reflectivity
Data Storage • Storage device: USB, in-built solid state hard disk (4 GB)
• Data file size: 3 GB
Input and Output

• I/O Ports: 2 USB, 1 mini USB and 1 Ethernet port
• Video out: Via VNC encoder: 1 or 2 axis quadrature
• Digital inputs 2 input lines (5 V TTL)
• Digital outputs 4 output lines (5 V TTL, 20 mA) for

  alarm or other external control
• Power output 5 V
• 350 mA current limited

Battery and Power Supply

• Battery type intelligent Li-ion
• Number of batteries 1
• Battery life typical: 7 hours in UT mode,

  6 hours in PA mode

Environmental Specifications • IP rating: Designed to meet IP66,
• Operating temperature -10 °C to 45 °C
  (14 °F to 113 °F)
   Storage temperature -25 °C to 60 °C
  (-13 °F to 140 °F)

 

Ultrasound 

 

    Conventional UT/TOFD Phased Array (PA)
General Connectors 4 x Lemo 1 or BNC IPEX
  Number of Focal Laws n/a 128
  Configuration 2 Channel 16:16 or 16:64
  Test Mode Pulse Echo, Transmit Receive and TOFD Pulse Echo, Transmit/Receive
Pulsers Pulse Voltage -100 V to -450 V (in steps of 10 V) -25 V to -75 V (in steps of 5 V)
  Pulse Width Adjustable: Spike to 2000 ns (2.5 ns resolution) Adjustable: Spike to 1000 ns (2.5 ns resolution)
  Pulse Shape Negative square wave (with ActiveEdge) Negative square wave (with ActiveEdge)
  Output impedance 5 Ω <10 Ω
Receivers Gain 100 dB (0.1 dB steps) Analogue gain 0 to 76 dB (0.1 dB steps) Analogue gain
  Input Impedance 1 kΩ (pitch and catch) 200 Ω
  System Bandwidth 200 kHz to 22 MHz (-3 dB) 200 kHz to 14 MHz
Data acquisition Scan Type A-Scan & TOFD S-Scan or L-Scan
  Number of scans Up to 2 1 (with up to 3 extracted A-Scans)
  Digitizing Frequency 50 MHz, 100 MHz, 200 MHz 65 MHz
  PRF 1 Hz to 1500 Hz 1 Hz to 5000 Hz
Data processing Max A-Scan Length 8192 samples 4096 samples
  Focussing Type n/a Natural, constant depth, constant path,
constant offset
  Rectifier Full wave, positive, negative, none (RF) Full wave, positive, negative, none (RF)
  Filtering

Analogue filters 4 (automatic or manual)

Digital filters 10 (automatic or manual)

Analogue filters 3 (automatic)

Digital filters 10 (automatic or manual)

Data Visualisation Cursor Types Cartesian, hyperbolic (TOFD) Cartesian, extraction box, angular
  Measurements Path length, depth, surface distance, DAC, AWS, DGS Path length, depth, surface distance, DAC, AWS
  Views A, B, C scan, 3D-view, Merged & TOFD A, B, C, L, S scan, 3D-view, Merged plus true top & end
  Number of layouts 18 35
TCG / DAC Number of points 16 16
  Maximum Slope 60 dB/μs 50 dB/μs
Alarms Number of Alarms (LED) 2 (sync on all gates & DACs) 2 (sync on all gates & DACs)
  Measurements (A Scan) Peak & flank (FSH, dB, depth, beam path length, surface distance), echo-to-echo,
floating gates (reference from IFT)
Peak & flank (FSH, dB, depth, beam path length, surface distance), echo-to-echo,
floating gates (reference from IFT)
Software Special features IFT, .csv data output, analysis software IFT, .csv data output, analysis software
  Report generation Pdf with embedded pdf reader, PNG screen capture Pdf with embedded pdf reader, PNG screen capture

 