https://www.trongkamol.com/product-page/trongkamol-starrett-mkp?lang=th
top of page
Proceq UT8000

Proceq UT8000

  • Brand: Proceq (Switzerland) 

 

  • The first wireless ultrasonic flaw detector

 

  • Connect the powerful base unit to your iPad and get to the real work faster

 

  • Intelligent software features

General

Proceq® UT8000 (793 10 500) Base unit, battery pack, batteries, power supply, belt holder, screwdriver, documentation, carrying case
Compliance EN 12668-1:2010; ASTM E317-16
User Interface Languages English, Spanish, French, German,
Japanese, Chinese, Korean
Transducer connections LEMO 00
Data storage Up to 1 terabyte (TB), depending on iPad
model
Battery type and life 6x AA rechargeable batteries; 5 h autonomy; compatible with off-the-shelf USB power banks
Power requirements Universal AC (100-240 V, 50-60 Hz)
Display type 2048 × 1536 / 326 PPI; fully laminated with antireflective coating and up to 120 Hz refresh rate
Display dimensions 7.9" – 12.9" diagonal
Overall Dimensions (W × H × D) 215 mm × 101 mm × 36 mm
8.46” x 3.97” x 1.42”
Weight 698 g, incl. battery pack
1.54 lbs incl. battery pack
USB ports 2x USB-C
Video output

Digital; wireless over Apple TV; HDMI/VGA/DisplayPort (via third-party adapter)

 

Environmental ratings

IP rating IP67
Shock tested MIL-STD-810F; Method 516.5; Procedure I, 6 cycles each axis; 15 g; 11 ms half sine
Vibration tested MIL-STD-810F; Method 514.5; Procedure I, Annex C; Figure 6; general exposure; 1 h / axis
Operating temperature Proceq UT8000 -10 to 50 °C (14 to 122 °F)
Battery operating temperature 0 to 40 °C (32 to 122 °F)
Battery storage temperature 0 to 50 °C (32 to 122 °F)

 

Pulser

Pulser Tunable square wave
PRF 10 – 2000 Hz, in increments of 10 Hz
Energy settings 50, 100 or 400 V
Pulse width Adjustable; 25 – 2500 nsec
Damping 50, 400 Ω

 

Receiver

Gain 0-110 dB
Receiver input impedance 400±5% Ω
Receiver bandwidth 0.25 – 20 MHz at -3 dB
EN12668 compliant
Digital filter settings 8 digital filter sets
Rectification Full-Wave; Positive Half-Wave; Negative Half-Wave; RF
System linearity Horizontal, ±0.5% FSW
Sampling frequency 125 MHz
Reject 0 – 100% FSH, in increments of 1%
Amplitude measurement 0.1% – 125% full screen height
Measurement rate Equivalent to PRF in all modes (single shot)

 

Calibration

Automated calibration Velocity, zero offset
Test modes Pulse echo, dual, through transmission or
grid
Units Millimeters, inches, or microseconds
Range 3 – 11760 mm at 5900 m/s
Velocity 100 – 10000 m/s
Zero offset 0 – 50 mm
Display delay up to full range
Refracted angle 0° – 90° in 1° increments

 

Gate

Measurement gates 2x, fully independent
Gate start Variable over entire displayed range
Gate width Variable, from 0.30 μs to end of displayed
range
Gate height Variable, 1% – 100% full screen height, in
increments of 1%
Alarms Positive and negative threshold/curve

 

Measurements

Measurement display locations Up to 5 locations (manual or auto selection)
Gate (1, 2) Thickness, sound path, projection, depth, amplitude, time-of-flight, min./max. depth, min./max. amplitude, sizing measurements based on mode
Echo-to-echo Standard gate 2 – gate 1
DAC/TCG Standard, up to 10 points, 110 dB dynamic
TCG range
Special DAC modes: Custom DAC (up to 6 curves),
20% – 80% view

 

Accessories 

Spare battery pack (793 30 110) Battery pack including 6xAA rechargeable
batteries
Tablet holder (793 30 120) iPad Mini – iPad Air
Supporting accessories Wide range of ultrasonic transducers, cables and calibration blocks, please call your local representative for more information.

 

bottom of page