Equotip UCI Portable Hardness Tester

SKU : Equotip UCI

Product Description

Equotip UCI Portable Hardness Tester

Product Description

  • แบรนด์ : Proceq (สวิตเซอร์แลนด์)
  • เครื่องวัดความแข็ง
  • เครื่องวัดความแข็งแบบพกพา

Product Features

Equotip 550 UCI

Flexible UCI hardness tester for fine-grained material with any shape and heat-treated surfaces. The patented adjustable test load enables a wide range of applications. Rugged touchscreen with enhanced software features and analysis functions.

Adjustable test load from HV1 to HV10

With this unique and patented feature a wide range of applications can be covered with out HV1-HV10 probe

Your Benefits

  • Combine with Leeb and Portable Rockwel
  • On-screen feedback to reduce measurement inaccuracies caused by the operator.
  • Ready-to-go reports through powerful built-in reporting feature

The UCI Measuring Principle

The UCI (Ultrasonic Contact Impedance) method uses the same pyramid-shaped diamond as a conventional Vickers hardness tester. Unlike Vickers testing, no optical evaluation of the indentation is required, enabling fast and portable measurements. The UCI method excites a rod into an ultrasonic oscillation. The test load is applied by a spring and typically ranges from 1 to 10 kg of force (HV1 HV10). As the diamond is forced into the material, the frequency of the rod oscillation changes in response to the contact area between the diamond and the material under test. The instrument detects the shift in frequency, converts it to a hardness value which is immediately displayed on the screen.


Product Technical

Equotip 550 Instrument
Display7 color display 800x480 pixels
MemoryInternal 8 GB Flash memory (up to 1000000 measurements)
Regional settingsMetric and Imperial units, multi-language and timezone supported
BatteryLithium Polymer, 3.6 V, 14.0 Ah
Battery lifetime> 8h (in standard operating mode)
Power input12 V +/-25 % / 1.5 A
Weight (of display device)About 1525 g (incl. Battery)
Dimensions250 x 162 x 62 mm
Max. altitude2500 m above sea level
Humidity< 95 % RH, non condensing
Operating temperature- 0 to 30°C (32 to 86°F) (Charging, running instrument)
- 0 to 40°C (32 to 104°F) (Charging, instrument is off)
- 10 to 50°C (14 to 122°F) (Non-charging)
EnvironmentSuitable for indoor & outdoor use
IP classificationIP 54
Pollution degree2
Installation category2


UCI Probe
Measuring range20 2000 HV
Resolution1 HV (UCI), 0.1 HRC
Measuring accuracy± 2 % (150 950 HV)
Test loads (in 10 N steps) SelectableHV1-HV10
Diamond indenterVickers diamond according to ISO 6507-2
Dimensions155 x Ø 40 mm (6.1 x 1.57 inches) without foot


Standardization  
Standards- ASTM A 1038
- ASTM E 140
- DIN 50159
- ISO 18265
- GB/T 34205-2017
Guidelines- ASME CRTD-91
- DGZfP Guideline MC 1
- VDI / VDE Guideline 2616 Paper 1
- Nordtest Technical Reports 424-1, 424-2, 424-3